Anaesthesia Intensive Care 1993, 21, 204 - 207

Laboratory Evaluation of the Go Medical PCI: A Disposable Patient-Controlled Analgesia Device

SUMMARY

A low-cost, disposable patient-controlled analgesia device was tested using a computor-controlled, automated testing station designed to simulate clinical use. Five devices were tested and delivered a mean bolus dose size of 95% of nominal, although one device achieved only 85% of nominal. Delivery was lowest immediately following an eight-hour interval during which demands were made. Further development and evaluation of this interesting device was warranted.